Basic Info.
Model NO.
NMYWX series
Warranty
1 Year
Transport Package
Carton & Wooden Packing
Trademark
NEWMEEN
Origin
China
HS Code
9031809090
Product Description
Applicable range:
The drug test chamber, with the scientific approaches, creates a good environment for the test of drug invalidity that needs long-term stability of temperature and humidity, it is the best selection for pharmaceutical enterprises to conduct drug stability test, applicable for drug long-term test, high-temperature test and high-light irradiation test, etc.
Product features:
1. The microcomputer controller and intelligent LCD are used to control and display the temperature and humidity respectively with the stability. accuracy and reliability.
2. Unique air circulating system ensures the uniform distribution of air in the working chamber.
3. The freezing machine adopts the internationally renowned brand original, full-enclosed compressor can switch automatically, to ensure long-term continuous operation of test equipment, with the features of stability, security and reliability.
4. The internal bladder is made of stainless steel and four semi-circular corners take convenience for easily cleaning.
5. RS-USB interface (optional)is available, which enables the incubator to be connected to the recorder or computer, to record the temperature and humidity for a long term.
6. A testing hole(optional) of diameter 25mm is available on the left of chamber body.
Conditions for stability test:
In the respects of functionality, performance and documents in Guidance of ICH,GMP and FDA defined the requirements. Meanwhile, Europe, Japan and USA had reached an agreement about a common stability test, this test is to collect the information, which is used for recommendation of raw material and drug stability, then to prove the drug validity under being exposed in some temperature, humidity, illumination or comprehensive environment.
*Storage conditions for the stability test of long-reserved sample:
Temperature: +25ºC ±2ºC
Humidity: 60 ±5%RH
Time: 12 months
Storage conditions for accelerating stability test:
Temperature: +40ºC ±2ºC
Humidity: 75 ±5%RH
Time: 6 months
Brightness under high-light irradiation condition: 4500 ±500LX
Technical parameters:
The drug test chamber, with the scientific approaches, creates a good environment for the test of drug invalidity that needs long-term stability of temperature and humidity, it is the best selection for pharmaceutical enterprises to conduct drug stability test, applicable for drug long-term test, high-temperature test and high-light irradiation test, etc.
Product features:
1. The microcomputer controller and intelligent LCD are used to control and display the temperature and humidity respectively with the stability. accuracy and reliability.
2. Unique air circulating system ensures the uniform distribution of air in the working chamber.
3. The freezing machine adopts the internationally renowned brand original, full-enclosed compressor can switch automatically, to ensure long-term continuous operation of test equipment, with the features of stability, security and reliability.
4. The internal bladder is made of stainless steel and four semi-circular corners take convenience for easily cleaning.
5. RS-USB interface (optional)is available, which enables the incubator to be connected to the recorder or computer, to record the temperature and humidity for a long term.
6. A testing hole(optional) of diameter 25mm is available on the left of chamber body.
Conditions for stability test:
In the respects of functionality, performance and documents in Guidance of ICH,GMP and FDA defined the requirements. Meanwhile, Europe, Japan and USA had reached an agreement about a common stability test, this test is to collect the information, which is used for recommendation of raw material and drug stability, then to prove the drug validity under being exposed in some temperature, humidity, illumination or comprehensive environment.
*Storage conditions for the stability test of long-reserved sample:
Temperature: +25ºC ±2ºC
Humidity: 60 ±5%RH
Time: 12 months
Storage conditions for accelerating stability test:
Temperature: +40ºC ±2ºC
Humidity: 75 ±5%RH
Time: 6 months
Brightness under high-light irradiation condition: 4500 ±500LX
Technical parameters:
Model | NMYWX-160C | NMYWX-300C | NMYWX-500C | NMYWX-1000C |
Volume | 160L | 300L | 500L | 1000L |
Ambient temp. | RT +5 -35ºC | |||
Temp. range | No illumination: 0-60ºC Illumination: 5-60ºC | |||
Intensity of illumination | 0-6000LX adjustable ±500LX | |||
Temp. resolution | 0.1ºC | |||
Temp. homogeneity | ±2ºC | |||
Temp. fluctuation | ±0.5ºC | |||
Humidity range | 20 ~95%RH | |||
Humidity fluctuation | ±3%RH | |||
Power supply | ~220V 50Hz. You can use voltage converters or customize according to regional differences. | |||
Temp. and humidity control mode | Balanced temperature and humidity control mode | |||
Refrigeration system /mode | Two sets of independent original brand full-enclosed compressor, automatic transfer in turn | |||
Controller | Microcomputer control | |||
Humidity sensor | Capacitive sensor | |||
Outline dim. (mm) | 550x450x680 | 610x510x1000 | 580x580x1480 | 1050x610x1655 |
Shelves (normal) | 3 | 3 | 4 | 4 |
Timing range | 1 ~9999min | |||
Remark | In a model, "A" represents the drug stability test chamber, "H" represents the drug high-light stability test chamber. and "C" represents the comprehensive drug stability test chamber. |